Repair of Mindray Brand Syringe Pump Type Benfusion SP1
Keywords:
Syrene Pump, Potentionmeter, Function TestAbstract
Syringe pump is a tool used to inject medicinal liquids into the human body with high precision using millimeters per hour (ml/h). The Mindray brand syringe pump type Benefusion SP1 had a fatal damage condition because the system experienced a system failure, so it could not function properly. The syringe pump cannot detect the syringe size that has been set because the control system does not work. Repair of the syringe pump was carried out by measuring using a multimeter at the specified measurement point. The syringe pump was successfully repaired by replacing the potentiometer component. After making recommendations, the flowrate and occlusion function tests were carried out. From the results of the flowrate function test, the percentage values were obtained, namely 10ml/h= 3.48%, 50ml/h= 1.19%, 100ml/h= 3.48%. and with occlusion experiments wiht a setting point of 100ml/h, an average value of 523mmHg or 10.113251 psi was easily obtained. From the results above it can be concluded that the tool is still suitable for use because it is still within the tolerance limit of 10% so that the tool can be used again.
References
Anggriani , R. (2019). FLOW RATE CONTROL SYSTEM ON SYRINGE PUMP . Bandar Lampung: Digilib Unila.
Baharudin , RN (2023). Repair of BBRAUN PERFUSOR COMPACT brand syringe pump . Semarang: Semarang Stikes.
Dimas . (2022). ACMA brand syringe pump repair type AF-330 . Semarang: Semarang Stikes.
Farid, M. (2023). Maerk syringe pump repair Mindray type Benefusion SP1 . Semarang: Semarang Stikes.
Hernawan , TY (2018). Repair of ACMA brand syringe pump type AF-330 . Semarang: Semarang Stikes.
Wisdom, NF (2013). Design build a syringe pump based Atmega8535 microcontroller. Journal Physics and Applied Science , 3-6.
Laia, A. (2022). SYRINGE PUMP CALIBRATION ANALYSIS . Journal PearlElectromedicine , 5.
Mutoharoh , C. (2022). ACMAFUSION AF-330 brand syringe pump repair . Semarang: Semarang Stikes.
Nur, B. (2018). Know More Details Regarding Syringe Pump Medical Devices. PT Glorya Medica Abadi, 5-7.
Paramartha , IK (2020). Repair Terumo brand syringe pump type STC-527 . Semarang: Semarang Stikes.